FAQ: Why am I getting a too many points error when I try to do a digitized measurement but I am not exceeding the specified number of points?

Oddly enough, this is most likely due to another module(s) in your mainframe that you are not programming. This is most likely due to the fact that you are using the COUNt command to take multiple measurements. The COUNt command applies to the entire mainframe. The default values for the number of points in the modules you are not programming is too high for the COUNt that you are using. Lower the number of points on the other modules in your mainframe.

Applies to: N3300A, N3301A, N3302A, N3303A, N3304A, N3305A, N3306A, N3307A

Select Home to return to the table of contents for more FAQs.